About the instrument

Our facility offers 2 CytoFLEX S instruments, equipped with

  • four lasers (488 nm, 561 nm, 638 nm und 405 nm) and overall 13 fluorescent channels 
  • two lasers (488 nm, 638 nm) and 6 fluorescent channels 

Samples can be taken from 96-well plates (U, V, flat bottom) as well as tubes (Eppendorf cups, 5 ml tubes). Additionally, the 2-laser instrument is able to operate on 96-deep-well plates. During the measurement in plates, it is possible to cool the plate loader. The cell concentration is detected without any additional counter beads.  

How does it work?

The cells or particles pass the laser in a sheath fluid with high velocity. According to the shape, size and/or staining of the cell or particle, scattered light is emitted, which is then detected by filters and WDMs (wavelength division multiplexers). Whereas the forward scatter signal depends on size, the side scatter signal is up to the complexity and granularity of the cell or particle. In addition, a fluorescent signal can be produced by staining of certain properties of cells or particles with dyes which directly bind, or with fluorescently labelled antibodies.